PARAMTM FT-F1 Fogging Tester is for high-temperature volatilizable constituent evaporation evaluation of the inner decoration components in automotives and aircrafts, including automotive inner plastic decoration components, polyurethane, textiles, leather, adhesives, non-woven and thermal forming elastomers, etc., as well as for fogging testing of automotive front high intensity discharge (HID) lamp.
Characteristics
6 test chamber design, simultaneous specimen test and blank test;
Stable operation, accurate data;
High precision temperature control.
Principle
The specimen is heated in the glass beakers and begins to evaporate; the volatile constituents condense on the glass plate or foil that has been cooled by the cooling chamber. After the cooling process, take off the glass plate or the foil. Through measuring the fogging value or weight of the condensed constituent on the glass plate or foil, and comparing with the data before condensation, the volatility of the specimen can be achieved.
3 Testing Methods
Gloss Method: the specimen is heated in the glass beaker; its evaporated constituents are condensed on the low temperature glass plate. Through the comparison and calculation of the gloss values before and after condensation, the fogging value of the specimen can be obtained.
Fogging Method: the specimen is heated in the glass beaker; its evaporated constituents are condensed on the low temperature glass plate. Through the comparison and calculation of the fogging values before and after condensation, the fogging value of the specimen can be obtained.
Weighing Method: the specimen is heated in the glass beaker; its evaporated constituents are condensed on the low temperature foil. Through calculating the changes of foil’s weight before and after condensation, the weight of the condensed constituents can be obtained.
Structure
PARAM FT-F1 Fogging Tester consists of high-temperature constant temperature bath, low-temperature constant bath, cooling plate, glass beaker, glass plate, reflectmeter, sample cutter and other accessories; this tester can accomplish sampling, heating, condensation and testing processes.
Operation Demo.
Prepare samples -- Switch on high-temperature and low-temperature constant bath -- Clean glass beakers and glass plates -- Place specimens -- Place glass plates or aluminum foils -- Place cooling blocks -- Specified time testing -- Get glass plates or aluminum foils down and wait until a specified time -- Measure reflectometer value of glass plates or weigh aluminum foils-- Compare and make conclusions
Technical Indexes
High Temperature Bath Range: room temperature~150ºC
(optional: room temperature~280ºC)
Resolution of High Temperature Bath: ±0.1ºC (150ºC)
Low Temperature Bath Range: 0ºC~100ºC
Resolution of Low Temperature Bath: ±0.1ºC
Dimensions of High-Temperature Bath: 670mm(L) x490mm(B)x 540mm(H)
Dimensions of Low-temperature Bath: 400mm(L)x 220mm(B)x 520mm(H)
Net Weight of High Temperature Bath: 32kg (heat-transfer medium exclusive)
Net weight of low temperature bath: 15kg (heat-transfer medium exclusive)
Power: AC 220V 50Hz
Standards
DIN 75201, ISO 6452, SAE J1756, QB/T 2728, BS EN 14288, PV 3920, PV 3015, ES-X83231, NES M0161, D45 1727, GM 9305P, TSM 0503G
Configuration
Standard: mainframe, constant temperature controller, sample presser, glass beakers, fluorine rubber sealing ring, metal ring, square glass plate, round glass plate, aluminum foil, aluminum foil sample cutter, lid, glass plate shelf, sample cutter, glossmeter, heat medium oil, DIDP, DOP, accessory shelf.
Optional: glass beakers, fluorine rubber sealing ring, square glass plate, round glass plate, aluminum foil, aluminum foil sample cutter, glass plate shelf, heat medium oil, glossmeter, electronic balance (0.01mg), DIDP, DOP, accessory shelf.